The Programat P100 allows for customized firing programs tailored to specific dental ceramic materials. Setting a Custom Program
: A membrane-sealed keypad for convenient data entry and program selection.
: Features standard programs for Ivoclar materials like IPS Classic and IPS d.SIGN, with additional customizable slots for other porcelain systems. ivoclar programat p100 manual english
The English manual includes factory-set programs for popular Ivoclar materials:
A: Go to Menu > Units > Temperature . The English manual confirms the P100 supports both, though Celsius is standard for dental ceramics. The Programat P100 allows for customized firing programs
Do not use water or chemical agents inside the chamber.
Connect to a grounded electrical outlet with the appropriate voltage. The furnace should not be used with extension cords. The English manual includes factory-set programs for popular
Users can switch between Celsius (°C) and Fahrenheit (°F) and select different buzzer signals through the internal menu. 3. Temperature Calibration (The Silver Test)
: It is critical to perform a silver wire calibration test regularly to ensure the furnace is firing at the correct temperature. Deviations can lead to under-fired (milky) or over-fired (slumped) restorations. Vacuum Check
The Programat P100 is specifically designed for firing dental ceramic materials. Using the furnace for other purposes (e.g., curing non-dental materials) can damage the equipment and void the warranty.